Dinamap Pro 1000 with Printer US version
Product Details:
Gold Standard Care for Any Situation
Anyone can detect normal blood pressure values. But where DINAMAP technology excels is in detecting the extremes, which are more difficult to determine and more critical to get right.
There's a reason BP, pulse, SpO2 and ECG are called vital signs. Blood pressure changes can be an early warning sign of many cardiac events. Pulse oximetry variance can be another. Apnea can take effect up to 15 seconds before an arrest. Adding a 3-lead Marquette ECG can also give you precious additional early warning response time. The DINAMAP Pro 1000 is a valuable patient assessment tool for unpredictable situations - rugged enough for transport even by air.
- The Pro 400 monitor sets the gold standard
for vitals: DINAMAP NIBP, Masimo SpO2 or Nellcor,
and IVAC TurboTemp
- DINAMAP Pro 100 - 400 monitors are covered
by a 3-year parts and labor warranty
- All DINAMAP monitors feature a common user interface
with GE monitors throughout the continuum of care, so
floating nurses can assess patients with minimal training
- DINAMAP NIBP technology is also available in our
Solar modular monitors and Dash configured monitors
Blood Pressure
- Blood Pressure Technique:
Oscillometric measurement of systolic, diastolic and mean arterial pressure and pulse rate.
- Cuff Inflation Adjustment Range:
Adult/Pediatric: 100 to 250 mmHg; 5 mmHg increments
Neonate: 100 to 140 mmHg; 5 mmHg increments
- Cuff Pressure Range:
Adult/Pediatric: 0 to 290 mmHg
Neonate: 0 to 140 mmHg
Automatically released if error condition is detected, power fails, or unit is switched off.
- Target Inflation Pressure Range:
User-selectable cuff inflation pressure
Adult/Pediatric: nominally 160 mmHG
Neonate: nominally 110 mmHg
- Determination Time:
20 seconds to 45 seconds typical
Adult - 120 secs maximum
Neonate - 85 secs maximum
Accuracy:
Meets or exceeds SP 10-1992 AAMI standards
- Pressure Preset Levels:
Resolutions of 5 mmHg up to Maximum Adult/Neonate limits
- Systolic Determinations:
Adult: Max: 245 mmHg; Min: 30 mmHg
Neonate: Max: 140 mmHg; Min: 40 mmHg
- Diastolic Determinations:
Adult: Max: 195 mmHg; Min: 10 mmHg
Neonate: Max: 100 mmHg; Min: 20 mmHg
- MAP Determinations:
Adult: Max: 215 mmHg; Min: 15 mmHg
Neonate: Max: 115 mmHg; Min: 30 mmHg
- Operating Modes:
AUTOMATIC, STAT, MANUAL Modes. Press START/STOP key to cancel.
- Automatic Mode (Blood Pressure) Cycle Time:
Preset at 3 minutes. User can select blood pressure determination frequency in varied increments between 1 and 120 minutes or operate by manual intervention. The result of the most recent determination will be displayed for up to 90 minutes (AUTO) or 2 minutes (MANUAL).
- Stat Mode:
Rapid BP determination with early systolic display within 5-minute period.
Pulse Rate Range:
Adult/Pediatric: 30 - 200 BPM
Neonate: 30 200 BPM
- Pulse Rate Accuracy:
BP±3.5%
- Overpressure Cut-off:
Adult/Pediatric: 300 - 330 mmHg
Neonate: 150 - 165 mmHg
Alarms
- System Alarm Functions:
Brief alarm test at power turn-on, alarm for no determination and a continuous alarm for microprocessor failure. Audible alarms with diagnostic codes displayed in pulse rate window for system failure.
- Patient Alarm Functions:
Systolic: Preset at 30 mmHg low and 180 mmHg high
Diastolic: Preset at 15 mmHg low and 130 mmHg high
- MAP:
Preset at 50 mmHg low and 140 mmHg high for adult/pediatric and neonate
- Pulse Rate:
Preset at 40 BPM low and 160 BPM high
- SpO2 Alarm Limits:
Preset at off (100%) high and 90% low.
Range: High 51-100% (100% = off)
Low 50-99%
- Alarm Indicators:
LED with audio alarm silence key. LED only flashes if audio alarm silenced and an alarm condition exists. Diagnostic data indicated on LCD.
Trend
- Prior Data:
Monitor stores 24 hours of data; up to 100 readings within time period.
Environmental
- Operating Temperatures:
+41°F to +104°F (+5°C to +40°C)
- Atmospheric Pressure:
700 to 1060 hPa
- Storage Temperature:
-4°F to +122°F (-20°C to +50°C)
- Humidity:
0 to 95% noncondensing
Power Requirements
- Input Power:
AC input voltage: 100-240 VAC, 50/60 Hz (nominal) 90-253 VAC, 47-63 Hz (range), 50VA. DC input voltage: 24 VDC, 36VA, supplied from a source conforming to IEC 601-1.
- Fuse:
AC input is protected by two internal fuses. DC input line is protected by an internal auto-resetting fuse. Battery: 12 volt, 2.3 amp-hours protected by internal auto-resetting fuse.
- Battery:
Sealed lead acid 12 volt 2.3 amp-hour, minimum operation time of 2 hours (5-minute auto cycle with adult cuff, temperature, SpO2 continuous and printer activated. Longer battery life with vital signs at less frequent intervals). Time for full recharge is 2 hours from full discharge with monitor switched off.
- Low Battery Warning:
Flashing LED and audio tone alert user of low battery condition.
- Power Cable:
Domestic: 10-ft, detachable, gray-jacketed, 16-gauge terminated with 3-prong hospital grade plug.
International: 7.5-ft (2.3m), detachable, black jacketed, 16-gauge terminated with appropriate plug.
**IVAC® Temperature (PRO 200 & PRO 400 only)
- Scale:
°Fahrenheit(F) °Celsius(C)
- RangePredictive mode:
Max: 42.2° C; 108.0°
Min: 31.6° C; 88.9° F
- Monitor Mode:
Max: 42.2° C; 108.0° F
Min: 26.7° C; 80.0° F
- Monitor Mode Accuracy:
±0.1°C ±0.2°F (when tested in a calibrated liquid bath; meets ASTM E1112, Table 1, in range specified)
- Predictive Mode Accuracy:
±1.0°F ±0.6°C Determination time less than 60 seconds
Use only IVAC probes and probe covers. The size, shape, and thermal characteristics of the probe covers can affect the performance of the instrument. Inaccurate readings or retention problems may occur unless IVAC probes and probe covers are used.
**IVAC is a trademark of Alaris Medical, Inc.
*NELLCOR™ Pulse Oximetry (SpO2) (PRO 300 & PRO 400 only)
Accuracy (± 1 Standard Deviation):
Adult/neonate 70 - 100% ±3.0 digits; under 70% unspecified
Saturation Range:
0 - 100%
Pulse Rate Range:
20 - 250 BPM
Pulse Rate Accuracy:
SpO2, ±3 BPM
Saturation Pitch:
Pitch changes with saturation. Volume selectable from off to 9.
*NELLCOR is a trademark of Mallinckrodt, Inc.
Physical Dimensions
- Dimensions:
9.8 in (H) x 9.8 in (W) x 6.9 in (D)
24.8 cm (H) x 24.8 cm (W) x 17.5 cm (D)
- Weight:
7.8 pounds; 3.5 kg (including battery)
- Mountings:
Self-supporting on rubber feet. Pole mountable. Carrying handle recessed into top of monitor.
- Portability:
Can be hand-carried by the recessed handle. When attached to the 3215 Rolling Stand, the monitor and accessories can be wheeled from patient to patient.
Printer
- Paper Width:
1.75 inches
- Writing Method:
Direct thermal
- Paper
Environmental requirements:
Temperature: 100°F (38°C) maximum
Humidity: 80% maximum
Related Products: