Corometrics 172 Basic

Corometrics 172 Basic
SKU: GE-Ol72WAT-B

This week only: FREE GROUND SHIPPING within continental
USA with your order of $475.00 or more! (applies only to online orders 20.00 lbs. or less)

$14,148.86
Product Description

170 Series Features assure uncompromising Antepartum Monitored Care.

  • Single Or Dual ultrasound versions available for non-stress testing and external monitoring.

  • Large LED displays offer at-a-glance confirmation of fetal heart rate and uterine activity from a distance.

  • High/Low heart rate alarm notifies clinician when high or low fetal heart rates exceed the user specified linits.

  • 9-Crystal Pulsed Doppler ultrasound technology provides a wide-beam profile for easier fetal heart rate detection.

  • Heart rate offset mode (dual ultrasound version) clearly seoerates twin fetal heart rates for easier visual assessment

PERFORMANCE SPECIFICATIONS

POWER REQUIREMENTS
Line Voltage: 90-260 VAC
Line Frequency: 50/60 Hz
Power Consumption: Less than or equal to 30 VA
Monitor DC Input: 12 Vdc at 2.5 A

PHYSICAL CHARACTERISTICS
Height: 5.75 in (14.6 cm)
Width: 16.75 in (42.5 cm)
Depth: 10.0 in (25.4 coat
Weight: <8 lbs (<4.5 kg)

ENVIRONMENTAL CONDITIONS
Monitor:
Ambient Temperature -
Operating: 50'F to 104'F (10'C to 40'C)
Storage: 14'F to BIT (-10'C to 55'C)
Relative Humidity -
Operating: 10% to 75%, non-condensing
Storage: 10% to 90%, non-condensing
Strip Chart Paper:
Ambient Temperature -
Operating: 50'F to 104'F (10'C to 40'C)
Storage: <80'F (<26.5'C)
Relative Humidity -
Operating: 30% to 70%, non-condensing
Storage: 45% to 65%, non-condensing

CERTIFICATION

UL 2601-1: Designed to meet UL-2601.1 Medical electrical equipment classified by Underwriter's Laboratories, Inc., with respect to fire, shock and mechanical hazards in accordance with UL-260 1. 1.

OPERATING MODES

Ultrasound Mode
Technique: Pulsed Doppler
with auto-correlation processing
Transducer Type: 9-crystal
Pulse Repetition Frequency: 2 kHz (all modes)
Pulse Duration: 92�s
Transmitter Frequency: 1.151 MHz
Intensity: <5mW/cm2(Isata)
Heart Rate Counting Range: 50-210 BPM
Leakage Current: Complies with IEC 601.1 and/or
IEC 601.1.1 harmonized national standard

STRIP CHART RECORDER
Heart Rate Scale
Scale 1 Scale 2
Chart Width: 7 con 8cm
Scaling: 30 BPM/cm 20 BPM/cm
Range: 30-240 BPM 50-210 BPM
Resolution: 1 BPM 1 BPM
Uterine Activity Scale
Tocotransducer
Chart Width: 4cm
Scaling: 25 relative units/cm
Range: 0-100 relative units
Resolution: 1 relative unit
Recorder Drive
Speeds: 1, 2 and 3 cm/min
Speed Accuracy: �2% over 10 mins